The allegation is against 1 of 2 leads; however, it is unknown which lead, therefore, all potential components are being listed.Additional components potentially involved in the event include: common device name: drg slim tip lead, model: mn10350-50a, udi: 05415067027139, serial: (b)(4), batch: 8754815.Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
|
It was reported that during procedure to pull patient's trial leads the physician had difficulty removing the right l3 lead.As such, once the lead tail came out, the lead tip with all four contacts remained in situ.X-rays showed the lead tip fragment was still intact and located by the spinous process.Surgical intervention may be undertaken to address this issue.
|