Model Number 37612
Device Problems Low impedance (2285); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2023
Event Type  malfunction  
Event Description
It was reported that the manufacturer representative rep said the healthcare provider hcp checked impedance in january, and low impedance was noted on electrodes 0, 1, and 2.The patient went in to check because of how quickly the battery depleted.The hcp programmed using electrode 3, and the patient got a benefit.Hcp checked impedance 2-3 weeks ago and today, but only electrode two was noted as having a short.Rep wanted to know how the other electrodes didn't appear as an issue.Technical services(ts) told rep the issue most likely didn't resolve but rather was hiding; ts recommended testing impedance in different head positions.It was said to the rep that as long as the patient was getting good therapy, there was no need for surgical revision; however, it was up to the hcp to decide.The caller was not with the patient.The patient was redirected to their healthcare provider to address the issue further.Additional information was received from the manufacturer representative (rep) that the surgeon was scheduling an exploratory/revi sion surgery.
 
Manufacturer Narrative
Concomitant medical products: product id 37085-60 lot# serial# (b)(4) implanted: (b)(6) 2010 product type extension product id neu_unknown_lead lot# serial# unknown implanted: (b)(6) 2010 product type lead product id 37085-60 lot# serial# (b)(4) implanted: (b)(6) 2010 product type extension.Other relevant device(s) are: product id: 37085-60, serial/lot #: ((b)(4), ubd: 09-apr-2014, udi#: (b)(4) ; product id: neu_unknown_lead, serial/lot #: unknown, ubd: 09-apr-2014, product id: 37085-60, serial/lot #: (b)(4), ubd: 05-feb-2014, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the manufacturer¿s representative (rep) reported the revision was not scheduled as of this time.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (M
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key16721523
MDR Text Key313111840
Report Number3004209178-2023-04857
Device Sequence Number1
Product Code MRU
UDI-Device Identifier00613994934611
UDI-Public00613994934611
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2013
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/03/2023
Initial Date FDA Received04/11/2023
Supplement Dates Manufacturer Received04/24/2023
Supplement Dates FDA Received04/27/2023
Date Device Manufactured10/04/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Outcome(s) Required Intervention;
Patient Age41 YR
Patient SexFemale
-
-