DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3013-45
Device Problem Moisture or Humidity Problem (2986)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/03/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use , the cs100 intra-aortic balloon pump (iabp) unit had too much water vapor in the iab pipeline, the spare equipment was replaced immediately.There was no personal injury reported.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate the issue.The fse stated that there was too much dust in the decondenser.So, in order to solve the issue, the fse cleaned the dust from the decondenser.All functional and safety checks were performed to meet factory specifications and the unit was returned to the customer and cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16724876
MDR Text Key313153716
Report Number2249723-2023-01903
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107790
UDI-Public10607567107790
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3013-45
Device Catalogue Number0998-00-3013-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2008
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-