The reported issue was confirmed.The root cause of the reported issue is due to a shorted t2.Device boots to an alarm 74.Tech al geibel installed test harness to complete decon.Tank harness was replaced.The arctic sun 6000 passed all performance testing, calibration, electrical safety tests and is functioning.Dhr is not required as this is event not an out-of-box failure and therefore is not manufacturing related.The complaint or reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|