MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN STAT; ARCTIC SUN DEVICE Back to Search Results
Model Number 60000000
Device Problems Circuit Failure (1089); Electrical Shorting (2926)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 03/23/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the arctic sun device boots to an alarm 74 (non-recoverable system error).Biomed discussed this was indicative of a shorted t2 (outlet control temperature) thermistor.
 
Event Description
It was reported that the arctic sun device boots to an alarm 74 (non-recoverable system error).Biomed discussed this was indicative of a shorted t2 (outlet control temperature) thermistor.
 
Manufacturer Narrative
The reported issue was confirmed.The root cause of the reported issue is due to a shorted t2.Device boots to an alarm 74.Tech al geibel installed test harness to complete decon.Tank harness was replaced.The arctic sun 6000 passed all performance testing, calibration, electrical safety tests and is functioning.Dhr is not required as this is event not an out-of-box failure and therefore is not manufacturing related.The complaint or reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN STAT
Type of Device
ARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16725405
MDR Text Key313146113
Report Number1018233-2023-02500
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741161513
UDI-Public(01)00801741161513
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60000000
Device Catalogue Number60000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-