MEDTRONIC HEART VALVES DIVISION EVOLUT FX VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number EVOLUTFX-29 |
Device Problems
Activation Failure (3270); Material Split, Cut or Torn (4008)
|
Patient Problems
Cusp Tear (2656); Valvular Insufficiency/ Regurgitation (4449); Aortic Valve Insufficiency/ Regurgitation (4450)
|
Event Date 04/06/2023 |
Event Type
Injury
|
Event Description
|
Medtronic received information that during implant of this transcatheter bioprosthetic valve, severe aortic insufficiency was observed after the post- balloon aortic valvuloplasty (bav).It was suspected that the bav tore a leaflet on the transcatheter valve.A 2nd valve was implanted within the 1st valve.Moderate paravalvular leak (pvl) was observed after deployment.A post-bav was performed, however, the pvl did not improve.No additional treatment was reported.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received that the reason for the initial post balloon aortic valvuloplasty (bav) was an expansion issue with the valve.The type of severe aortic insufficiency that occurred was noted to be central.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H6 ¿ method, result and conclusion codes h10 ¿ conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.No valve implantation procedure images were available for medtronic review, and the valve remained implanted, so no product analysis could be performed.Without imaging, we are unable to conclusively determine the root cause for the reported expansion issues experienced and if there were any patient anatomy issues that may have prevented the valve from fully expanding.Several factors can contribute to the onset and propagation of structural valve dysfunction (including cuspal tear) such as patient medical history (age, disease stage, comorbidities, etc.), pharmacological factors, and/or intrinsic properties of the valve itself.In this case, it was suspected that the balloon aortic valvuloplasty (bav) tore a leaflet on the transcatheter valve.However, this could not be confirmed with the information available.Aortic regurgitation can be caused by a variety of factors, including valve positioning, patient anatomy, or presence of pre-existing patient conditions.A conclusive root cause of the regurgitation could not be determined, but the initial valve implant with the torn leaflet was a likely contributing factor.This event does not indicate device misuse or malfunction.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received that a 26 millimeter (mm) semi compliant non-medtronic balloon was used for the post balloon aortic valvuloplasty (bav).One inflation was performed.The balloon was expanded and then immediately deflated.
|
|