It was reported that the right ventricular (rv) lead had a rise to high threshold measurements.It was noted that the rv lead had an associated drop in r-waves and a slight drop in impedance measurements.It was further reported that the physician stated the rv lead was pulled back from the implant position per chest x-ray and the cardiac resynchronization therapy defibrillator (crt-d) may be lower on chest.During lead revision, upon opening the pocket, the lead was twisted and tangled.The physician suspected twiddlers or similar issue.It was noted that during the revision procedure, the rv lead impedance was out of range (oor).The lead was repositioned and a antibacterial absorbable envelope was placed to help stabilize the device.Both the crt-d and lead and remain in use.The rv lead threshold measurements returned to baseline.No further patient complications have been reported as a result of this event.
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|