MEDTRONIC EUROPE SARL ACTIVA; Stimulator, electrical, implanted, for parkinsonian tremor
|
Back to Search Results |
|
Model Number 37601 |
Device Problem
High impedance (1291)
|
Patient Problems
Ambulation Difficulties (2544); Unspecified Nervous System Problem (4426)
|
Event Date 06/02/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A supplemental report will be issued.
|
|
Event Description
|
It was reported that there was high impedance of 40,000 caused by/related to the ins.The patient had walking disorder, dystonia, and akinesia.The device was reprogrammed to use contact 2 in place of contact 3 and the issue was resolved.
|
|
Search Alerts/Recalls
|
|
|