MAUDE Adverse Event Report: DIASORIN INC. DIASORIN LIASON XL HSV 1 AND 2 TYPE SPECIFIC ANTIBODY TEST; ENZYME LINKED IMMUNOSORBENT ASSAY, HERPES SIMPLEX VIRUS, HSV-2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DIASORIN INC. DIASORIN LIASON XL HSV 1 AND 2 TYPE SPECIFIC ANTIBODY TEST; ENZYME LINKED IMMUNOSORBENT ASSAY, HERPES SIMPLEX VIRUS, HSV-2 Back to Search Results
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Insufficient Information (4580)
Event Date 01/18/2022
Event Type  Injury  
Event Description
Patient had five hsv 2 positive tests on the diasorin hsv 1 and 2 igg test, (index values = 0.9, 0.94, 1.15, 1.39, and 1.56), followed up with a western blot which was negative for hsv 2.Date given is the western blot confirmatory test.Reference reports mw5116642, mw5116643, mw5116644, mw5116645.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIASORIN LIASON XL HSV 1 AND 2 TYPE SPECIFIC ANTIBODY TEST
Type of Device
ENZYME LINKED IMMUNOSORBENT ASSAY, HERPES SIMPLEX VIRUS, HSV-2
Manufacturer (Section D)
DIASORIN INC.
MDR Report Key16750879
MDR Text Key313527139
Report NumberMW5116646
Device Sequence Number1
Product Code MYF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age26 YR
Patient SexFemale
Patient Weight59 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-