• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH PSI SD800.402 PEEK IMPLANT; PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH PSI SD800.402 PEEK IMPLANT; PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE Back to Search Results
Catalog Number SD800.402
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2023
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Reporter is a j&j employee.Complainant part is not expected to be returned for manufacturer review/investigation.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from depuy synthes reports an event in (b)(6) as follows: it was reported that on (b)(6), 2023 the customer reported the size of implant was planned too large, so the result was not satisfactory and had to be replaced.This report is for one (1) psi sd800.402 peek implant.This is report 1 of 1 for complaint pc b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot.Product code: sd800.402.Lot number: 4415p57 (me23-emo-kul).Manufacturing date: 02-feb- 2023.Manufacturing site: jabil mezzovico.No ncr's generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was not returned to depuy synthes for evaluation.The r&d team conducted and investigation based off the case records and complaint documentation.Per the complaint description above, the psi case files and communication were reviewed.The investigation included an end-to-end process review of the documentation and forms along with the surgeon complaint report and the received details on the case.No images intra or post operatively were received with the complaint.The implant design was completed and verified as per the depuy synthes design instructions and roles.After the surgeon reviewed the design and approved it with his signature, the device was produced and inspected according to process and released upon acceptable quality inspection.As per the communication stated from the surgeon above.The originally planned and designed device was reviewed and approved by the surgeon.However, post-operatively the cosmetic outcome was deemed not satisfactory, because the initial device was implanted and revision surgery was required, this case falls under the complaint process.Based on the information and documents available, this design investigation for the initial device is closed as not valid regarding a design related root cause.As a part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for the psi 60*20*40 peek.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PSI SD800.402 PEEK IMPLANT
Type of Device
PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK MEZZOVICO (CH)
via cavazz 5
mezzovico CO 6805
SZ   6805
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16753550
MDR Text Key313730741
Report Number8030965-2023-04749
Device Sequence Number1
Product Code GXN
UDI-Device Identifier07611819337749
UDI-Public(01)07611819337749
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K053199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSD800.402
Device Lot Number4415P57
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2023
Initial Date FDA Received04/17/2023
Supplement Dates Manufacturer Received04/27/2023
12/04/2023
Supplement Dates FDA Received05/12/2023
12/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/02/2023
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age36 YR
Patient SexFemale
Patient Weight70 KG
-
-