Model Number HL-90
Device Problem Defective Alarm (1014)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
B3: date of event and d4: udi number is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported the warmer interlock alarm was not working.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
One device was received.A visual inspection noted a cracked enclosure and broken light emitting diodes (led)'s.Filled tank with water, attached temperature check, plugged in line cord, and turned on the power switch for functional testing.The micro switch is not triggering the alarm the way it should confirming the complaint.The root cause was due to a bent microswitch lever from usage.A device history record (dhr) review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service in the previous twelve (12) months and there was no indication that the complaint was related to a previous service of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16753926
MDR Text Key313444201
Report Number3012307300-2023-04214
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No