The reported events of backup operation, ble (bluetooth) telemetry issue, and an error message were confirmed.Final analysis found that the device was in backup vvi and in this state, ble telemetry is not accessible.When interrogated as received on the bench, an error message was observed.Analysis of device image found the device went into backup mode due to memory error.After the device was restored from backup mode, functional testing was performed.Telemetry (ble and inductive), impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.The backup operation could not be reproduced.The cause of the reported event could not be determined.
|