SMITHS MEDICAL ASD, INC. SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7349-24
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problems Pain (1994); Inadequate Pain Relief (2388)
Event Date 03/14/2023
Event Type  malfunction  
Event Description
It was reported the disposable and pump were switched out as a concern for under delivery.The patient was not getting comfortable with epidural despite multiple patient-controlled boluses on top of the infusion.The epidural was disconnected from the pump.The pump and tubing were switched out and infusion continued; followed up with patient an hour later, epidural worked with no concerns.
 
Manufacturer Narrative
Other, other text: a supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other, other text: no product was returned.We are unable to confirm the reported complaint.If the product is returned, smiths medical will reopen this complaint for further investigation.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product was returned; therefore, no visual and functional tests were performed, the reported complaint could not be confirmed, and the root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET, ADMINISTRATION, INTRAVASCULAR
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16762240
MDR Text Key313865070
Report Number3012307300-2023-04255
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7349-24
Device Catalogue Number21-7349-24
Device Lot Number4339160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age32 YR
Patient SexFemale
-
-