Model Number 0998-00-0800-53 |
Device Problem
Backflow (1064)
|
Patient Problems
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/07/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a service repair by a getinge field service engineer, the cardiosave intra-aortic balloon pump (iabp) was found to have old dried blood in the pim and safety disk.There was no patient involvement.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
The getinge field service engineer (fse) that encountered the reported issue during during repair of pump and found old dried blood in the pim and safety disk.Replaced all blood contaminated parts.Performed a full functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released and cleared for clinical service.
|
|