Model Number CI512
Device Problem No Device Output (1435)
Patient Problem Failure of Implant (1924)
Event Date 03/31/2023
Event Type  malfunction  
Event Description
Per the clinic, the patient experienced subsequent loss of connection to the internal device.Hardware exchange and troubleshooting attempts were made; however, the issue could not be resolved.The implanted device remains.It is unknown if there are plans to explant the device and re-implant the patient with another cochlear device as of the date of this report.Additional information has been requested but it has not been made available as of the date of this report.
 
Manufacturer Narrative
This report is submitted on april 20, 2023.
 
Manufacturer Narrative
Per the clinic, the device was explanted on (b)(6) 2023, and the patient was re-implanted with another cochlear device during the same surgery.This report is submitted on october 27, 2023.
 
Manufacturer Narrative
Device analysis indicated device failure.Device analysis report attached.This report is submitted on march 21, 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI512 COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
ashwinii chelvan
unit ug-1, vertical podium
no. 8 jalan kerinchi
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key16774124
MDR Text Key313635739
Report Number6000034-2023-01246
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502014564
UDI-Public(01)09321502014564(11)110211(17)130210
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/10/2013
Device Model NumberCI512
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received02/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient Age59 YR
Patient SexFemale
-
-