It was reported that the bd maxplus¿ pressure rated extension set with needleless connector had a loose connection to the angio catheter, and sprayed out medication onto the patient's back as a result.This complaint was created to capture the 1st of 3 related incidents.The following information was provided by the initial reporter: "the manifolds have been coming apart and leaking, and the connection with the extention to the angio seems to not screw on very tight and leaks at that spot.Bd max plus pressure rated ext set - # mp5312-c not fully securing to angio cath where circled below causing leakage.Details: pt.Was in endo being put to sleep and the anesthesia provider stated that the connection of item 7592897 to the angio was leaking.He tightened that connection then went to the manifold (7570763), started giving propofol and the connection there came apart completely and sprayed the patient down her back.There was no adverse event.It just delayed the patient getting to sleep for the procedure.Procedure: egd and colonoscopy".
|
D.4.Medical device expiration date: unknown.H.4.Device manufacture date: unknown.H.6.Investigation summary: a complaint of connection to angio not connecting properly and leaking was received from the customer.A photo of the component with connection issues was provided.The defect cannot be seen in the photo.A device history record review could not be performed on model mp5312-c because the lot number is unknown.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10.
|