• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST TRADITION PRO HANDPIECE; HANDPIECE, AIR-POWERED, DENTAL PRODUCT CODE EFB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST TRADITION PRO HANDPIECE; HANDPIECE, AIR-POWERED, DENTAL PRODUCT CODE EFB Back to Search Results
Catalog Number 770445
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
This report summarizes 6 malfunction events where midwest tradition pro handpiece would not hold burs.No injury resulted.
 
Manufacturer Narrative
Additional serial numbers included in this report: (b)(6).6 of 6 devices were returned for evaluation.Evaluation of five devices found them all to be within specification.Evaluation of one device found chuck wear and lack of maintenance.The devices were cleaned of debris and the turbines were replaced.The devices were repaired and returned to the customers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST TRADITION PRO HANDPIECE
Type of Device
HANDPIECE, AIR-POWERED, DENTAL PRODUCT CODE EFB
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
york, PA 17401
7178494593
MDR Report Key16781397
MDR Text Key313935902
Report Number9614977-2023-00009
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K131319
Number of Events Reported6
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Catalogue Number770445
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/20/2023
Patient Sequence Number1
-
-