• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDHFA500Q
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Wound Dehiscence (1154); Unspecified Infection (1930)
Event Date 04/04/2023
Event Type  Injury  
Event Description
It was reported that a patient presented with an infection noted on the patient's implantable cardiovert defibrillator.The device was explanted and replaced on (b)(6) 2023.The patient was stable throughout.
 
Event Description
New information received notes that the patient exhibited wound dehiscence on the patient's device.It was unknown if cultures of the pocket were taken.
 
Manufacturer Narrative
A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.Based on this information, the device was found to communicate appropriately with a merlin programmer and has not reached the elective replacement indicator (eri).The cause of infection could not be traced to the device.Correction: b1: field should be marked adverse event only.Correction: b2: field should reflect adverse event of intervention and infection adverse event.Correction: h1: field should be marked "serious injury".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16781462
MDR Text Key313713680
Report Number2017865-2023-18073
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberP000154400
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/04/2023
Initial Date FDA Received04/20/2023
Supplement Dates Manufacturer Received04/20/2023
05/24/2023
Supplement Dates FDA Received05/09/2023
05/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE LEADS.; QUARTETLEADS.; RIGHT ATRIAL LEAD.
Patient Outcome(s) Required Intervention; Other;
Patient Age85 YR
Patient SexMale
-
-