• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS VASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS VASCULAR STENT GRAFT Back to Search Results
Catalog Number FVM09060
Device Problems Positioning Failure (1158); Material Perforation (2205); Misfire (2532); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2023
Event Type  malfunction  
Event Description
It was reported that during a stent graft placement procedure in external iliac artery for occlusion of iliac trunk, the stent graft allegedly failed to deploy due to high friction in the deployment system.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the stent delivery system was returned for evaluation.The stent graft was found partially deployed and a stent strut was found perforating and deforming both the marker band and the outer sheath.The perforation of the sheath by the strut is assessed as being the main issue, which led to partial deployment of the stent graft.It was reported that a 0.035" guidewire/9fr introducer were used, the device was flushed, the lesion was calcified / slightly tortuous, and the device was placed in the straight segment of the lumen.Based on the provided information and the evaluation of the returned sample, the investigation is confirmed for outer sheath perforation, partial deployment, and deformation.A definite root cause for the reported event could not be determined.Labeling/packaging review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding the preparation of the device the instructions for use states "flush the stent graft lumen with sterile saline using a small volume syringe.Attach the syringe to the luer-lock port of the y-injection-adapter on top of the delivery system and apply positive pressure.Ensure that the 2-way stopcock is open also flush saline through the inner catheter lumen via the luer-lock port at the back of the delivery system.Flush until saline leaks from the tip of the catheter and all air is removed".Regarding the precautions prior and during deployment the instructions for use states "prior to stent graft deployment in tortuous anatomy, ensure that the proximal stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible failure to deploy".The instructions for use states regarding the accessories "a super stiff guide wire (0.035 in.) is advanced from a femoral artery puncture site.Use an introducer sheath for the implant procedure".H10: d4 (expiry date: 02/2025).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS VASCULAR STENT GRAFT
Type of Device
VASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16782824
MDR Text Key313832327
Report Number9681442-2023-00148
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741145681
UDI-Public(01)00801741145681
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
P130029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFVM09060
Device Lot NumberANGP2576
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/23/2023
Initial Date FDA Received04/21/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-