Model Number SUPPORT ARM 177 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/27/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the support arm for patient circuit broke.There was no patient involvement.Manufacturer's ref #: (b)(4).
|
|
Manufacturer Narrative
|
It was reported that the support arm bracket broke.Our field service engineer investigated the ventilator on site and solved the issue by replacing the support arm bracket.Regular functional tests were carried out and the failure did not cause any injuries to patients or operators or delays in procedures.A broken support arm may lead to to stop of ventilation (extubation) or injury.The root cause as to why the support arm bracket broke cannot be established by this investigation.
|
|
Event Description
|
Manufacturers ref: #(b)(4).
|
|