• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE CONTINENTAL EUROPEAN 230 VOLT LEVEL 1 HOTLINE; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE CONTINENTAL EUROPEAN 230 VOLT LEVEL 1 HOTLINE; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Overheating of Device (1437)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2023
Event Type  malfunction  
Manufacturer Narrative
Date of event is unknown, no information has been provided to date.No 510k as product not sold in us.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the device was overheating "above 43 degrees".Patient involvement unknown.
 
Manufacturer Narrative
Other, other text: additional information: b3, b5, and h6 - health effects codes: updated.
 
Event Description
Additional information was received confirming the event occurred during the preparation phase.No patient harm, injury, or adverse effects.
 
Manufacturer Narrative
One device was received.Upon removing the front enclosure light emitting diode (led) fell out and was otherwise in good physical condition.Powered the device on and let run for an hour and it worked properly.Temperature was stable on 41.9 degrees celsius and after pressing the "over temperature alarm test" button the device stopped, but alarm did not sound that could have been caused by the broken light emitting diode (led).Unable to replicate or confirm the complaint.A device history record (dhr) review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service in the previous 12 months and there was no indication that the complaint was related to a previous service of the device.The printed wiring assembly (pwa) was replaced.No further action was taken as the complaint could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTINENTAL EUROPEAN 230 VOLT LEVEL 1 HOTLINE
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16790434
MDR Text Key314009845
Report Number3012307300-2023-04381
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberHL-90-UK-230
Was Device Available for Evaluation? No
Date Returned to Manufacturer04/12/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-