Model Number HL-90 |
Device Problem
Overheating of Device (1437)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/31/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event is unknown, no information has been provided to date.No 510k as product not sold in us.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the device was overheating "above 43 degrees".Patient involvement unknown.
|
|
Manufacturer Narrative
|
Other, other text: additional information: b3, b5, and h6 - health effects codes: updated.
|
|
Event Description
|
Additional information was received confirming the event occurred during the preparation phase.No patient harm, injury, or adverse effects.
|
|
Manufacturer Narrative
|
One device was received.Upon removing the front enclosure light emitting diode (led) fell out and was otherwise in good physical condition.Powered the device on and let run for an hour and it worked properly.Temperature was stable on 41.9 degrees celsius and after pressing the "over temperature alarm test" button the device stopped, but alarm did not sound that could have been caused by the broken light emitting diode (led).Unable to replicate or confirm the complaint.A device history record (dhr) review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service in the previous 12 months and there was no indication that the complaint was related to a previous service of the device.The printed wiring assembly (pwa) was replaced.No further action was taken as the complaint could not be confirmed.
|
|