BECTON, DICKINSON & CO., (BD) BD VACUTAINER® MULTIPLE SAMPLE LUER ADAPTER; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367300
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2023
Event Type  malfunction  
Manufacturer Narrative
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that during use with bd vacutainer® multiple sample luer adapter the adapter broke in half.This occurred on 2 occasions.There was no report of patient impact.The following information was provided by the initial reporter, translated from french to english: luer adapters are used to connect the trasis carpule (containing fdg) to the patient's infusion.At the time of injection, a luer adapter split in two.The activity is lost as the capsule is unusable resulting in a delay of the injection and the monopolisation of a lead tank (injection device) during the decay time.This problem has already been encountered twice with the same batch in this department.
 
Manufacturer Narrative
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by torque testing and no issues were observed relating to breaks apart as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode breaks apart.Bd was not able to identify a root cause for the indicated failure mode.H3 other text : see h.10.
 
Event Description
It was reported that during use with bd vacutainer® multiple sample luer adapter the adapter broke in half.This occurred on 2 occasions.There was no report of patient impact.The following information was provided by the initial reporter, translated from french to english: luer adapters are used to connect the trasis carpule (containing fdg) to the patient's infusion.At the time of injection, a luer adapter split in two.The activity is lost as the capsule is unusable resulting in a delay of the injection and the monopolisation of a lead tank (injection device) during the decay time.This problem has already been encountered twice with the same batch in this department.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® MULTIPLE SAMPLE LUER ADAPTER
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16790684
MDR Text Key313775987
Report Number1024879-2023-00247
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K991088
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number367300
Device Lot Number2315779
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-