Catalog Number UNKTRAPEASE
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2023
Event Type  malfunction  
Manufacturer Narrative
As reported, a trapeze inferior vena cava filter was found to be fractured.There was no reported patient injury.The device is not available for evaluation as this is a permanent filter it will be left in place.The product is unavailable for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease ivc filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.The instructions for use (ifu) states filter fracture is a potential complication of vena cava filters.Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut.Without imaging available for review and the limited information provided, the reported event could not be confirmed or further clarified.There is nothing in the information provided to suggest that the reported event is related to the design and/or manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported, a trapeze inferior vena cava filter was found to be fractured.There was no reported patient injury.Additional information was requested but not available.The device is not available for evaluation as this is a permanent filter it will be left in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEASE
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
santiago troncoso 808
juarez, chihuahua 32574
MX   32574
MDR Report Key16792798
MDR Text Key313814791
Report Number9616099-2023-06471
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKTRAPEASE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK.
Patient Age62 YR
Patient SexFemale
Patient Weight83 KG
Patient EthnicityHispanic
-
-