• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PROVISIONAL STEMMED TIBIAL COMPONENT SIZE 4; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PROVISIONAL STEMMED TIBIAL COMPONENT SIZE 4; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/03/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that staff opened a tray and found a broken tibial trial inside.The surgical technique was utilized.There was no surgical delay.All available information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified sign of use and fractured.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.The root cause of the reported issue is attributed to wear and tear from repeated use.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROVISIONAL STEMMED TIBIAL COMPONENT SIZE 4
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16805899
MDR Text Key313904008
Report Number0001822565-2023-01094
Device Sequence Number1
Product Code HWT
UDI-Device Identifier00889024217713
UDI-Public(01)00889024217713(10)63581253
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00598103702
Device Lot Number63581253
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/03/2023
Initial Date FDA Received04/25/2023
Supplement Dates Manufacturer Received06/29/2023
Supplement Dates FDA Received06/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexMale
-
-