Model Number 21-7106-24 |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/11/2023 |
Event Type
malfunction
|
Event Description
|
It was reported the disposable has torn for the second time at the screw thread.Through the cracks, the medication ran along the line.No patient injury reported.
|
|
Manufacturer Narrative
|
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Device evaluation: one sample was returned in used condition.The visual inspection found no damage.During the functional testing, no leaks were identified.No alarms were activated when the pump was set to run.The customer reported failure was unable to be confirmed.During the dhr review, no relevant findings were found during the manufacturing process.
|
|