It was reported that there was an issue with pl574t - challenger ti-p sm-ligat.Clips 12 cartr.According to the complaint description, the surgeon used two clips for product testing before using the product, but the clips completely detached from the body and fell inside the patient's abdomen.The surgeon removed the clip that fell into the abdomen and equipped a new clip to the product.An additional medical intervention was necessary.Additional information was not provided nor available.The adverse event is filed under aag reference (b)(4).
|
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 2(5) x probability 1(5) of occurrence) according to din en iso 14971 is still acceptable.Explanation and rationale: unfortunately due to a lack of data and without the product we cannot determine the exact root cause for the mentioned deviation.According to the quality standard, a production error and a material defect can most probably be excluded.If further investigations are required, the product should be provided for examination.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
|