• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.; HANDHELD PRODUCTS & LIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.; HANDHELD PRODUCTS & LIGATION Back to Search Results
Model Number PL574T
Device Problem Unintended Ejection (1234)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2023
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with pl574t - challenger ti-p sm-ligat.Clips 12 cartr.According to the complaint description, the surgeon used two clips for product testing before using the product, but the clips completely detached from the body and fell inside the patient's abdomen.The surgeon removed the clip that fell into the abdomen and equipped a new clip to the product.An additional medical intervention was necessary.Additional information was not provided nor available.The adverse event is filed under aag reference (b)(4).
 
Manufacturer Narrative
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 2(5) x probability 1(5) of occurrence) according to din en iso 14971 is still acceptable.Explanation and rationale: unfortunately due to a lack of data and without the product we cannot determine the exact root cause for the mentioned deviation.According to the quality standard, a production error and a material defect can most probably be excluded.If further investigations are required, the product should be provided for examination.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.
Type of Device
HANDHELD PRODUCTS & LIGATION
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key16807512
MDR Text Key313927516
Report Number9610612-2023-00084
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL574T
Device Catalogue NumberPL574T
Device Lot Number52770258
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/13/2023
Initial Date FDA Received04/25/2023
Supplement Dates Manufacturer Received05/04/2023
Supplement Dates FDA Received05/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-