Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2022
Event Type  Injury  
Event Description
Patient had a hsv 1 positive test (index value = 2.71) and a hsv 2 positive test (index value = 3.23) on the diasorin hsv 1 and 2 igg test, followed up with a western blot which was negative for hsv 1 and 2.Date given is the western blot confirmatory test.Reference report #mw5116988.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIAISON HSV-2 TYPE SPECIFIC IGG
Type of Device
ENZYME LINKED IMMUNOSORBENT ASSAY, HERPES SIMPLEX VIRUS, HSV-2
Manufacturer (Section D)
DIASORIN INC.
MDR Report Key16808273
MDR Text Key314073630
Report NumberMW5116989
Device Sequence Number1
Product Code MYF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/24/2023
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age25 YR
Patient SexMale
Patient Weight73 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-