Model Number 106
Device Problems Erratic or Intermittent Display (1182); Premature Elective Replacement Indicator (1483)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2022
Event Type  malfunction  
Event Description
It was reported that the physician had disabled the patient & #39;s generator in order to evaluate the vns therapy.Prior to the disablement, the generator was observed to have low battery.When the generator was re-enabled, the generator & #39;s battery was observed at a higher reading.No other relevant information has been received to date.
 
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova & #39;s employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any & #34;defects¿ or & #34;malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
Tablet data was provided for review.From review of the data, the cause for the fluctuating battery status is due to the battery voltage being near the 11-25% threshold, and therefore there is no indication of a device failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key16809023
MDR Text Key313958779
Report Number1644487-2023-00510
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public0105425025750061
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/22/2022
Device Model Number106
Device Lot Number205187
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 03/31/2023
Initial Date FDA Received04/25/2023
Supplement Dates Manufacturer Received01/23/2024
Supplement Dates FDA Received02/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/04/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age26 YR
-
-