• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-DETEK INCORPORATED PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE; DEFIB/PACING ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-DETEK INCORPORATED PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE; DEFIB/PACING ELECTRODE Back to Search Results
Model Number 8900-4006
Device Problem Arcing of Electrodes (2289)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2023
Event Type  malfunction  
Event Description
Complainant alleged that while attempting to cardiovert a 65-year-old male patient, an arc was seen from the electrode pads.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Manufacturer Narrative
The pads were returned to zoll medical corporation for evaluation without the original packaging opened.The customer's report was not replicated or confirmed.The pads were stuck to a napkin and were unable to be separated for any testing.A visual inspection of the returned electrode pads found hair and discoloration around the tins of the pads.The pads were scrapped after testing.The instructions for use (ifu) for the pro-padz electrodes provides instructions on proper skin preparation and electrode application.The ifu also warns: excessive hair can inhibit good coupling (contact), which can lead to the possibility of arcing and skin burns; poor adherence and/or air under the electrodes can lead to the possibility of arcing and skin burns.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRO-PADZ RADIOLUCENT SOLID GEL ELECTRODE, SINGLE
Type of Device
DEFIB/PACING ELECTRODE
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
Manufacturer Contact
tina lombari
525 narragansett park drive
pawtucket, RI 02861
4017291400
MDR Report Key16809365
MDR Text Key313947570
Report Number1218058-2023-00089
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946016661
UDI-Public00847946016661
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8900-4006
Device Catalogue Number8900-4006
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/04/2023
Initial Date FDA Received04/25/2023
Supplement Dates Manufacturer Received04/04/2023
Supplement Dates FDA Received06/12/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
-
-