BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; IMPLANTABLE DEFIBRILLATOR Back to Search Results
Model Number G125
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); High Capture Threshold (3266)
Patient Problem Asystole (4442)
Event Date 02/12/2023
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited noise, oversensing and high pacing thresholds on the right ventricular channels.Further evaluation of the patient was discussed.This device remains in service.No further adverse patient effects were reported.Additional information was received detailing that the left ventricular lead exhibited pacing inhibition.It was decided to continue monitoring the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM CRT-D
Type of Device
IMPLANTABLE DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16820243
MDR Text Key314082363
Report Number2124215-2023-20202
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589126
UDI-Public00802526589126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/06/2020
Device Model NumberG125
Device Catalogue NumberG125
Device Lot Number130919
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/06/2023
Initial Date FDA Received04/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-