BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G156 |
Device Problems
Premature Discharge of Battery (1057); Failure to Charge (1085); Incorrect, Inadequate or Imprecise Result or Readings (1535); Delayed Charge Time (2586); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/31/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.This crt-d also recorded a code 1007 indicative of a charge time greater than 45 seconds.In addition, a battery capacity depleted message was displayed.This crt-d remains in service at this time, however device replacement was recommended.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
The local area field representative was contacted for additional information.If information is provided in the future, a supplemental report will be issued.
|
|