Model Number 5403190
Device Problems Difficult to Flush (1251); Suction Problem (2170); Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2023
Event Type  malfunction  
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 04/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that six days post dialysis catheter placement, the catheter allegedly clotted.It was further reported that the catheter was removed and replaced.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one 19cm pristine d/l catheter was returned for evaluation.In addition to the returned physical sample, two electronic photos were provided for review.Functional, tactile, gross visual and microscopic visual evaluations were performed.The investigation is inconclusive for the reported catheter clotted issue, suction problem and difficult to flush, as the exact circumstances at the time of the reported event was unknown and the issue cannot be reproduced in the lab.A definitive root cause could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiration date: 04/2024), h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that six days post dialysis catheter placement, the catheter allegedly clotted.It was further reported that the catheter did not work.Reportedly, the catheter was removed and replaced.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISTINE LONG-TERM HEMODIALYSIS CATHETER
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16826704
MDR Text Key314125961
Report Number3006260740-2023-01577
Device Sequence Number1
Product Code MSD
UDI-Device Identifier00801741219030
UDI-Public(01)00801741219030
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5403190
Device Catalogue Number5403190
Device Lot NumberREGT4594
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/20/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age42 YR
Patient SexFemale
Patient Weight90 KG
Patient RaceBlack Or African American
-
-