• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. TRACHEAL INTUBATION FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. TRACHEAL INTUBATION FIBERSCOPE Back to Search Results
Model Number LF-DP
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, when cleaning the tracheal intubation fiberscope with endoscope reprocessor, the device was improperly and insufficiently reprocessed and cleaned by connecting a modified maj-829-leak test air tube for the endoscope reprocessor.The issue was found during reprocessing.There were no reports of patient harm.Related patient identifiers: (b)(6).
 
Manufacturer Narrative
The device was not returned to olympus for evaluation.A supplemental report will be submitted if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reprocessing issue was the facility staff did not know that the modified oer-2 connecting tube should not be used for oer-3 for reprocessing.There is a possibility that the modified connecting tube for qer-2 was used for oer-3 and reprocessed.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRACHEAL INTUBATION FIBERSCOPE
Type of Device
TRACHEAL INTUBATION FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16826948
MDR Text Key314128138
Report Number9610595-2023-06820
Device Sequence Number1
Product Code EOQ
UDI-Device Identifier04953170340192
UDI-Public04953170340192
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K981543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLF-DP
Device Lot Number2137570
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/29/2023
Initial Date FDA Received04/27/2023
Supplement Dates Manufacturer Received05/31/2023
Supplement Dates FDA Received06/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/26/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MAJ-829 (MODIFIED PRODUCT); OER-3
-
-