(b)(4).Visual inspection of the returned device confirms that the device was fractured and had marks consistent with usage.The device history records were reviewed and no discrepancies were identified.The device had been in the field for approximately seven years and six months with unknown usage.However, based on the information available, the root cause was determined to be normal wear and tear during the instrument¿s field life.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|