• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; CHAIR, POSITIONING, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; CHAIR, POSITIONING, ELECTRIC Back to Search Results
Model Number PLR3955LT
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Burn(s) (1757)
Event Date 04/20/2023
Event Type  Injury  
Manufacturer Narrative
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
 
Event Description
The provider alleges the customer alleges that the lift chair has two burn spots on it and the lift chair allegedly burnt his back.
 
Event Description
The provider alleges the customer alleges that the lift chair has two burn spots on it and the lift chair allegedly burnt his back.
 
Manufacturer Narrative
There were 2 corresponding burn marks at the seat base interface.Given the uni-directional discoloration of the poly-fill underneath the burn spots along with the normal operating temperature of both heat pads, the evaluator believes the burn spots were caused by an outside heat source unrelated to the chair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Type of Device
CHAIR, POSITIONING, ELECTRIC
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york avenue
duryea, PA 18642
8008008586
MDR Report Key16830882
MDR Text Key314160190
Report Number2530130-2023-00034
Device Sequence Number1
Product Code INO
UDI-Device Identifier00606509401500
UDI-Public00606509401500
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070950
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberPLR3955LT
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 04/20/2023
Initial Date FDA Received04/28/2023
Supplement Dates Manufacturer Received04/20/2023
Supplement Dates FDA Received04/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/03/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-