Model Number PLR3955LT |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
Burn(s) (1757)
|
Event Date 04/20/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
|
|
Event Description
|
The provider alleges the customer alleges that the lift chair has two burn spots on it and the lift chair allegedly burnt his back.
|
|
Event Description
|
The provider alleges the customer alleges that the lift chair has two burn spots on it and the lift chair allegedly burnt his back.
|
|
Manufacturer Narrative
|
There were 2 corresponding burn marks at the seat base interface.Given the uni-directional discoloration of the poly-fill underneath the burn spots along with the normal operating temperature of both heat pads, the evaluator believes the burn spots were caused by an outside heat source unrelated to the chair.
|
|