Model Number 21-7301-24 |
Device Problems
Device Alarm System (1012); Product Quality Problem (1506)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/17/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other, other text: a supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported the disposable caused the pump to alarm no disposable, clamp tubing".The patient started using the cassette on apr/13.Then, on apr/17, a "no disposable, clamp tubing" alarm went off.The patient detached the cassette, raised the tubing position slightly, and attached the cassette to the pump again.After that, the alarm was settled.No patient injury reported.
|
|
Manufacturer Narrative
|
Other, other text: device evaluated, picture one shows cassette product out of its original packaging.Picture two shows the front view of a cassette.Picture three shows the top view of cassette.Picture four shows the pressure plate latch section of the cassette.Visual inspection showed the sample was received without the blue clip.No damages, kinks, cuts or any other damage detected.Functional testing performed and the failure mode could not be duplicated by functional testing, complaint is not confirmed.No root cause determined due to complaint was not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
|
|