Model Number 21-7301-24
Device Problems Device Alarm System (1012); Product Quality Problem (1506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/17/2023
Event Type  malfunction  
Manufacturer Narrative
Other, other text: a supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported the disposable caused the pump to alarm no disposable, clamp tubing".The patient started using the cassette on apr/13.Then, on apr/17, a "no disposable, clamp tubing" alarm went off.The patient detached the cassette, raised the tubing position slightly, and attached the cassette to the pump again.After that, the alarm was settled.No patient injury reported.
 
Manufacturer Narrative
Other, other text: device evaluated, picture one shows cassette product out of its original packaging.Picture two shows the front view of a cassette.Picture three shows the top view of cassette.Picture four shows the pressure plate latch section of the cassette.Visual inspection showed the sample was received without the blue clip.No damages, kinks, cuts or any other damage detected.Functional testing performed and the failure mode could not be duplicated by functional testing, complaint is not confirmed.No root cause determined due to complaint was not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD CASSETTE RESERVOIRS WITH FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16831430
MDR Text Key314169547
Report Number3012307300-2023-04713
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027215
UDI-Public10610586027215
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number21-7301-24
Device Catalogue Number21-7301-24
Device Lot Number4354548
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-