Model Number G138 |
Device Problems
Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
|
Patient Problem
Asystole (4442)
|
Event Date 01/01/2022 |
Event Type
Injury
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillators (crt-d) device exhibited noise and oversensing on the right ventricular (rv) channel.This oversensing of the noisy signals led to pacing inhibition.The healthcare professional was able to reproduce the noise with isometrics, however oversensing was not seen.There were concerns for possible asystole greater than 2 seconds.Boston scientific representative indicated that the patient was seen in clinic and were able to reproduce the noise with isometrics but there was no oversensing observed.Patient was using an electric tiller potentially at the time of the possible organized emi type noise.There was no change in impedance during the noise.This crt-d device remains in service.No patient adverse effects were reported.Additional information received indicated that the tiller was not the cause of the oversensing.The patient is going to have the rv lead revision in upcoming months.No adverse effects were reported.
|
|
Manufacturer Narrative
|
This report captures the explant of this device and impact on the patient.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillators (crt-d) device exhibited noise and oversensing on the right ventricular (rv) channel.This oversensing of the noisy signals led to pacing inhibition.The healthcare professional was able to reproduce the noise with isometrics, however oversensing was not seen.There were concerns for possible asystole greater than 2 seconds.Boston scientific representative indicated that the patient was seen in clinic and were able to reproduce the noise with isometrics but there was no oversensing observed.Patient was using an electric tiller potentially at the time of the possible organized emi type noise.There was no change in impedance during the noise.This crt-d device remains in service.No patient adverse effects were reported.Additional information received indicated that the tiller was not the cause of the oversensing.The patient is going to have the rv lead revision in upcoming months.No adverse effects were reported.Additional information received indicated that the device was explanted and successfully replaced.No additional patient adverse effects were reported.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillators (crt-d) device exhibited noise and oversensing on the right ventricular (rv) channel.This oversensing of the noisy signals led to pacing inhibition.The healthcare professional was able to reproduce the noise with isometrics, however oversensing was not seen.There were concerns for possible asystole greater than 2 seconds.Boston scientific representative indicated that the patient was seen in clinic and were able to reproduce the noise with isometrics but there was no oversensing observed.Patient was using an electric tiller potentially at the time of the possible organized emi type noise.There was no change in impedance during the noise.This crt-d device remains in service.No patient adverse effects were reported.Additional information received indicated that the tiller was not the cause of the oversensing.The patient is going to have the rv lead revision in upcoming months.No adverse effects were reported.Additional information received indicated that the device was explanted and successfully replaced.No additional patient adverse effects were reported.
|
|
Manufacturer Narrative
|
The returned device was inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.All setscrews moved freely in both directions.Dimensional analysis of the header was completed.A manual shock test was performed on the device.And the device passed the test at 42-ohms.All five seal plugs on the header were inspected under high power and a hole was found on the lv seal plug.A hole on the lv seal plug would not cause noise on the rv channel.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.This report captures the explant of this device and impact on the patient.
|
|
Search Alerts/Recalls
|