• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE Back to Search Results
Model Number M00564760
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2023
Event Type  Injury  
Event Description
It was reported to boston scientific corporation on april 7, 2023, that an ultraflex tracheobronchial covered distal stent was to be implanted in the left principal bronchus to treat a malignant stricture due to a tumor during an airway stent implantation procedure performed on (b)(6) 2023.During the procedure, the 10mm x 40mm ultraflex tracheobronchial covered stent was successfully deployed; however, under bronchoscope, it was found that a stent wire was fractured near the bulge region.The ultraflex tracheobronchial stent was removed from the patient with forceps and a 10mm x 30mm laminated stent was used to complete the procedure.There were no reported patient complications as a result of this event.The patient's condition following the procedure was reported to be stable.
 
Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of stent break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16835599
MDR Text Key314262760
Report Number3005099803-2023-02235
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K012883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 04/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/08/2023
Device Model NumberM00564760
Device Catalogue Number6476
Device Lot Number0027103521
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/07/2023
Initial Date FDA Received04/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight60 KG
-
-