BD MEDICAL (BD WEST) MEDICAL SURGICAL BD¿ PRE-FILLED NORMAL SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH
|
Back to Search Results |
|
Catalog Number 306593 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the bd¿ pre-filled normal saline syringe stopper was damaged when blood was withdrawn.The following information was provided by the initial reporter, translated from chinese; "when the indwelling needle is given to the patient, the rubber stopper in the needle barrel is broken when the blood is withdrawn, and it is replaced again.".
|
|
Manufacturer Narrative
|
H.6.Investigation summary: it was reported the rubber stopper is broken when the blood is withdrawn.As a sample was not returned, a thorough sample investigation could not be completed.It could be possible the customer is not using the product as intended.This product is designed to flush the iv line and not for drawing blood.A device history record review was completed for provided material number 306593, lot 2040574.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.To date, there have been no other similar events reported for this lot.Based on the investigation and with no sample analysis the symptom reported by the customer could not be confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Search Alerts/Recalls
|
|
|