U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER Back to Search Results
Model Number D97130F5
Device Problem Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that a d97130f5 swan pacing catheter from lot 64493592 had a failure of no capture.The steps taken by the customer were changing the cables first, which did not correct the problem.The solution was to change the catheter and it worked properly.They are no longer in possession of the catheter.This occurrence was during an interventional case.There was no reported harm to patient.
 
Manufacturer Narrative
Without the return of the product, it is not possible to determine if damages or defects exist on the product, nor can any manufacturing nonconformance, failure mode, root cause, or potential contributing factors be identified.No actions will be taken at this time.A device history record review was initiated to assess that the device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER
Type of Device
SWAN-GANZ BIPOLAR PACING
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500294
MDR Report Key16843665
MDR Text Key314279459
Report Number2015691-2023-12675
Device Sequence Number1
Product Code DYG
UDI-Device Identifier00690103150438
UDI-Public(01)00690103150438(17)240803(11)220804(10)64493592
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K822723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD97130F5
Device Catalogue NumberD97130F5
Device Lot Number64493592
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/10/2023
Initial Date FDA Received05/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-