Model Number 4FC12 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226); Pericardial Effusion (3271)
|
Event Date 04/05/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, when exchanging sheaths, the wire positioned in the left atrium was lost.An attempt was made to rewire into the left atrium and advance the sheath across the septum.Imaging confirmed the desired position was not achieved.The patient became hypotensive.Cardiac tamponade and pericardial effusion were confirmed and a pericardiocentesis was carried out.The case was aborted under general anesthesia.The patient was transferred to another hospital for open heart surgery. the patient's hospitalization was extended.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: a photo was returned and analyzed.After reviewing the picture, no anomaly was found on the sheath.In conclu sion, the reported "desired position was not achieved" issue due to the sheath was not confirmed through data analysis.The physical product was not returned.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|