Model Number IPN915182
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2023
Event Type  malfunction  
Event Description
It was reported that a clip could not be loaded into the applier properly during use.The user opened another cartridge and could load the clips without problem to complete the procedure.No clip fell/remained in the patient.No patient harm or injury.Patient status reported as "fine".
 
Manufacturer Narrative
Qn#(b)(4).
 
Manufacturer Narrative
(b)(4).The customer returned one cartridge from one unit of 544230 hemolok ml clips 6/cart 84/box for investigation.The returned sample was visually examined with and without magnification.Three broken clips were returned in the cartridge.One hook half and three pierced boss halves were discovered within the returned cartridge.No other defects or anomalies were observed.Visual examination revealed that the broken clips were broken in half at the hinge.The clips breaking at the hinge during loading was determined to be the result of insert mismatch at the hinge area of the clip.Functional inspection could not be performed as there were no intact clips remaining in the cartridge.A device history record review was performed and no relevant findings were identified.The ifu for this product states, "always check the alignment of the applier jaws before use.When closed, jaw tips should be directly aligned and not offset.Alignment of the jaw is critical for safe application of the clip.If this is not done, patient injury may occur.Proper maintenance, care and cleaning are necessary to ensure proper functionality." the reported complaint was confirmed based upon the sample received.Three broken clips were returned in the cartridge.One hook half and three pierced boss halves were discovered within the returned cartridge.Visual inspection confirmed that the clips were broken at the hinge.The clips breaking at the hinge during loading was determined to be the result of insert mismatch at the hinge area on the pierce leg side of the hinge.The root cause is due to inadequate specifications for insert mismatch in the highest stress area of the hinge, which is a design related issue.A capa has been previously opened to further investigate issues related to clip breakages.Teleflex will continue to monitor and trend on complaints of this nature.Other remarks: n/a.Corrected data: n/a.
 
Event Description
It was reported that a clip could not be loaded into the applier properly during use.The user opened another cartridge and could load the clips without problem to complete the procedure.No clip fell/remained in the patient.No patient harm or injury.Patient status reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOLOK ML CLIPS 6/CART 84/BOX
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16856134
MDR Text Key314466731
Report Number3003898360-2023-00756
Device Sequence Number1
Product Code FZP
UDI-Device Identifier24026704695915
UDI-Public24026704695915
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K133202
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN915182
Device Catalogue Number544230
Device Lot Number73B2200435
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/13/2023
Initial Date FDA Received05/02/2023
Supplement Dates Manufacturer Received05/30/2023
Supplement Dates FDA Received06/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-