SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number CA32K0/300/004JP |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the product was observed.No patient injury reported.
|
|
Manufacturer Narrative
|
D4 udi is unknown.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during this device history record review.A product sample was received for evaluation.Visual and functional testing were performed.A hole was found in the serpentine ridge about 60 cm from the tip of the patient connector.A leak test revealed a slight air leak.Five (5) pictures were attached, during the analysis conducted it could be observed in one sample a with focus on section with a tier damaged.The root cause of the reported issue was found to be undetermined.No corrective actions were taken since the investigation did not reveal that the defect was caused during the manufacturing process.
|
|