Model Number 383547
Device Problem Premature End-of-Life Indicator (1480)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2023
Event Type  malfunction  
Event Description
Device explanted due to eos.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Manufacturer Narrative
Upon receipt, the device interrogation revealed the battery status eos.A number of 29 charging cycles was documented.The amount of charge taken from the battery was verified and found to be normal and as expected.The eos status was removed with a technical programmer and the ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied and the device detected the fibrillation signal as specified and started the charging of a defibrillation shock, however the charging was aborted, indicating a depleted battery.In conclusion, the icd was fully functional.The battery status eos was anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILESTO 7 HF-T DF-1
Type of Device
CRT-D
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key16861600
MDR Text Key314479076
Report Number1028232-2023-02223
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 08/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2015
Device Model Number383547
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age62 YR