Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.Detailed analysis did not nonspecific product performance allegation as baseline testing completed on the device found no failing conditions.All testing that was completed on the device passed or was not applicable, therefore no problem was detected.Also, muscle stimulation allegation could not be confirmed by lab analysis moreover, known inherent risk was based on the field report of muscle stimulation.Muscle stimulation is a known medical risk for implanted pulse generator systems (pacemakers, defibrillators, rhythm therapy devices and associated cardiac leads).This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
|