Model Number V5
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Bowel Burn (1756)
Event Date 04/20/2023
Event Type  Injury  
Event Description
It was reported that on (b)(6), a novasure procedure was performed and a first device was used but they kept getting vacuum alarms and the width dial was filling up with blood.They used a second novasure device where the procedure went as normal, with no issues.The doctor scoped the patient after the procedure and the ablation looked good.Two days after the procedure on (b)(6) 2023 the patient went to the hospital complaining of pain where a laparoscopic diagnostic was done.It was confirmed that had a bowel injury.They said it looked like it burned through the wall and not a hole.The patient was taken to surgery and is still in the hospital.No additional information available.
 
Manufacturer Narrative
Hologic is submitting this report in accordance with 21.Cfr part 803.The submission is based upon the currently available information.The submission of this report does not represent a confirmation of device malfunction involving the hologic products in the event described.D4: lot and serial number of the device not provided by the complainant; therefore, the udi, expiration and manufacturing dates are not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.2 devices were involved in this case: 1222780-2023-00153.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVASURE V5
Type of Device
DEVICE, THERMAL ABLATION, ENDOMETRIAL
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
ariel lafuente
562 parkway
coyol free zone building b24
san jose 20102
MDR Report Key16864810
MDR Text Key314508456
Report Number1222780-2023-00152
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P010013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberV5
Device Catalogue NumberNSV5KITUS-001
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/26/2023
Initial Date FDA Received05/03/2023
Supplement Dates Manufacturer Received04/26/2023
Supplement Dates FDA Received05/15/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-