Model Number 6260-9-236
Device Problems Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Patient Problems Ambulation Difficulties (2544); Insufficient Information (4580)
Event Date 03/27/2023
Event Type  Injury  
Event Description
It was reported: symax stem with worn conus and loose head.
 
Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.H3 other text : not returned.
 
Event Description
It was reported: sym ax stem with worn conus and loose head.
 
Manufacturer Narrative
An event regarding disassociation involving an metal head was reported.The event was confirmed through analysis of the device.Method & results: product evaluation and results: visual inspection: visual inspection of the returned devices indicated that the trunnion of the stem is severely worn.The metal head appears to have damage from the disassociation event and explantation.Damage consistent with the loss of taper lock was observed on the head and stem of the returned devices.This has been documented as part of a capa.No further material analyses were performed.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the subject device has been identified to be within scope of an nc and capa.Lot specific voluntary recall ra 2016-028 was initiated for the lfit v40 cocr heads within scope of said nc and capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analyses identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 36MM/+5
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16865935
MDR Text Key314525702
Report Number0002249697-2023-00473
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
K022077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 07/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2012
Device Model Number6260-9-236
Device Catalogue Number6260-9-236
Device Lot NumberDH3MED
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2249697-08/29/2016-007R
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
-
-