Model Number WA53000A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, the 10mm rigid telescope, hd, quick lock, was producing a blurry or foggy image.Inspection and testing of the returned device found the eyepiece was damaged.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation.But the customer's allegation could not be confirmed.Additionally, the device evaluation found signs of damage and corrosion of the outer tube.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation (h6/h10).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be conclusively determined.It is likely the reported event (damaged eyepiece) occurred due to the use of excessive force from improper handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
RIGID OPTICAL LAPAROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16872815
MDR Text Key314641413
Report Number9610773-2023-01216
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052757
UDI-Public04042761052757
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K950076
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53000A
Device Catalogue NumberWA53000A
Device Lot Number671636
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/05/2023
Initial Date FDA Received05/04/2023
Supplement Dates Manufacturer Received07/26/2023
Supplement Dates FDA Received08/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-