• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC. SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC. SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 397002-001
Device Problems Battery Problem (2885); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/24/2023
Event Type  malfunction  
Manufacturer Narrative
The driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
 
Event Description
The customer, a syncardia certified hospital, reported the companion 2 driver received an emergency battery error.
 
Manufacturer Narrative
Alarm history and patient data file review found review found new alarms recorded in the companion 2 driver's data file: on apr 24th there was an emergency battery error alarm that occurred multiple times.Visual inspection of external components revealed no abnormalities.Visual inspection of internal components revealed a damaged capacitor on the power management board though unrelated to this complaint.This finding handled per standard non-conformance procedures.Companion 2 driver passed all incoming functional testing.The battery was then subjected to a ten-minute battery discharge test.During this testing the battery performed as intended.Failure investigation for this complaint confirmed the reported issue via alarm history data review.The complaint was not replicated via functional/observational testing.The root cause of the customer reported emergency battery alarm could not be conclusively determined.Failure investigation identified no test failure, damage, or abnormalities that could have contributed to the customer reported alarm.Driver functioned as designed.Patient was switched to a back-up driver with no reported adverse impact.This issue weill be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.(if new or additional information is received in the future, syncardia will file a follow-up mdr.) (b)(4).Follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key16872941
MDR Text Key314909743
Report Number3003761017-2023-00066
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/24/2023
Initial Date FDA Received05/04/2023
Supplement Dates Manufacturer Received04/24/2023
Supplement Dates FDA Received07/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-