• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; IMPLANTABLE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; IMPLANTABLE DEVICE Back to Search Results
Model Number G124
Device Problems Premature Discharge of Battery (1057); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/15/2023
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient has complete heart block felt it whenever there is threshold testing.Boston scientific technical services (ts) discussed that the device has reached battery capacity depleted (bcd) and uncertain if device has exhibited premature battery depletion (pbd) unless device is returned to the laboratory for testing and recommended immediate replacement.The device remains in service.No adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
 
Event Description
It was reported that the patient has complete heart block felt it whenever there is threshold testing.Boston scientific technical services (ts) discussed that the device has reached battery capacity depleted (bcd) and uncertain if device has exhibited premature battery depletion (pbd) unless device is returned to the laboratory for testing and recommended immediate replacement.The device remains in service.No adverse patient effects were reported.Additional information received indicating that the device was explanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.The returned cardiac resynchronization therapy defibrillator (crt-d) was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, defibrillation, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
 
Event Description
It was reported that the patient has complete heart block felt it whenever there is threshold testing.Boston scientific technical services (ts) discussed that the device has reached battery capacity depleted (bcd) and uncertain if device has exhibited premature battery depletion (pbd) unless device is returned to the laboratory for testing and recommended immediate replacement.The device remains in service.No adverse patient effects were reported.Additional information received indicating that the device was explanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM CRT-D
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16873514
MDR Text Key314636028
Report Number2124215-2023-21908
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589102
UDI-Public00802526589102
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/14/2019
Device Model NumberG124
Device Catalogue NumberG124
Device Lot Number487722
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/28/2023
Initial Date FDA Received05/04/2023
Supplement Dates Manufacturer Received04/28/2023
03/29/2024
Supplement Dates FDA Received05/09/2023
04/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age48 YR
Patient SexMale
-
-