• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-L 15 FR X 23 CM ANTE; CATHETER, HEMODIALYSIS, IMPLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-L 15 FR X 23 CM ANTE; CATHETER, HEMODIALYSIS, IMPLA Back to Search Results
Catalog Number CS-15232-VFIE
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem Insufficient Information (4580)
Event Date 04/21/2023
Event Type  malfunction  
Manufacturer Narrative
Qn#(b)(4).
 
Event Description
It was reported that: the cvc extensions leave a dent/kink at site of clamp affecting/reducing blood flow.The position of the clamp was changed, and the areas were massaged.No patient injury was reported but the patient's current condition is unknown.Additional information has been requested from the account and will be submitted with the follow up report.
 
Manufacturer Narrative
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.The ifu provided with this kit warns the user, " examine catheter and extension lines before and after each treatment for any signs of damage." a device history record review was performed based on a potential lot from sales history and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event Description
It was reported that: the cvc extensions leave a dent/kink at site of clamp affecting/reducing blood flow.The position of the clamp was changed, and the areas were massaged.No patient injury was reported but the patient's current condition is unknown.Additional information has been requested from the account and will be submitted with the follow up report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-L 15 FR X 23 CM ANTE
Type of Device
CATHETER, HEMODIALYSIS, IMPLA
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16875679
MDR Text Key314916782
Report Number9680794-2023-00340
Device Sequence Number1
Product Code MSD
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K141051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCS-15232-VFIE
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED; NOT REPORTED
-
-